Comparison of closed loop and sensorless adaptive optics in widefield optical microscopy
1
Department of Physics, Durham University, South Road, Durham, DH1 3LE, UK
2
Biophysical Sciences Institute, Durham University, South Road, DH1 3LE, UK
Received:
14
January
2013
Revised:
22
February
2013
We report on a closed loop widefield adaptive optics, optical microscopy system in which the feedback signal is provided by backscattered light from the sample acting as a guide star. The improvement in imaging performance is compared to an adaptive optics system controlled via an image optimisation routine commonly described as sensorless adaptive optics. The samples viewed were imaged without fluorescence to ensure that photobleaching and other potential variations did not affect the comparisons in system performance though the method is equally applicable for fluorescence microscopy. The closed loop system is self-optimising for different areas of the sample, using a common reference wavefront, with the accuracy of the loop being limited by variation across the sub-aperture images induced by guide star elongation. Optimisation using an image sharpness metric gives slightly sharper images but takes significantly longer. We thus believe that both wavefront sensor based closed loop AO and metric based optimisation have a role to play in AO for microscopy and that the method of backscattered light as a guide star has a great potential in the application of AO, particularly to optical coherence tomography.
Key words: Optical microscopy / adaptive optics / widefield microscopy / sensorless adaptive optics / closed loop adaptive optics
© The Author(s) 2013. All rights reserved.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.